In our previous article, we discussed titration accuracy for automatic Kjeldahl nitrogen analyzers and mentioned that detector sensitivity exerts an indirect influence over titration accuracy. This article focuses on the color detector in detail.
Color‑Judgment Criteria in National Standard 
According to the Kjeldahl method described in GB 5009.5‑2025, indicator colour serves as the criterion for identifying the titration endpoint, corresponding to purplish‑red / light grey‑red colours. With conventional manual Kjeldahl set‑ups, operators visually judge colour shifts throughout titration. While subjective judgment varies slightly among individuals, measurement error remains minimal as long as consistent judging criteria are applied for both blank tests and sample tests. For automatic Kjeldahl nitrogen analyzers, a built‑in color detector identifies titration endpoints automatically.
Working Principle of the Color Detector
Red, green and blue are the three primary colours; all other hues are generated by mixing these three components in different proportions. The national‑standard‑specified indicators are Methyl Red, Bromocresol Green and Methylene Blue
During titration for automatic Kjeldahl analysis, the instrument continuously monitors real‑time changes in the three primary‑colour signal values.
- For mixed indicator of Methyl Red + Bromocresol Green: Blue‑component variation does not occur. The instrument only tracks the ratio shift between red and green signals. The purplish‑red endpoint specified in the national standard corresponds to a fixed green‑to‑red signal ratio, which reproduces exactly the colour perceived by human eyes in manual titration.
- For mixed indicator of Methyl Red + Methylene Blue: Green‑component variation is absent. The instrument monitors only the ratio between red and blue signals. The light grey‑red endpoint defined by the standard matches a fixed blue‑to‑red signal ratio.
First‑Generation Color Detector (Filter‑Based Design)
This design utilises optical‑filter technology. A daylight lamp or LED illuminates liquid inside the titration cell from above. Two or four optical filters are mounted underneath. The system measures light intensity after white light passes through titrant liquid and the optical filters. Collected electrical signals are amplified and transmitted to the instrument mainboard, commanding the titrator to dispense standard acid and hold the target endpoint colour.
Symmetrical four‑quadrant color detectors are commonly deployed to enhance colour resolution and sensitivity. First‑generation detectors feature low hardware cost yet suffer prominent drawbacks: performance is highly susceptible to light‑source brightness, physical position and illumination angle. Detector position adjustment is mandatory every time after cleaning the titration cell to guarantee uniform light intensity across all four sensing elements, resulting in cumbersome operation. Colour drift may also occur, requiring regular inspection and calibration.
Second‑Generation Color Detector (CMOS‑Integrated Chip)
Driven by advances in integrated‑circuit and chip technologies, second‑generation color detectors adopt CMOS architecture. The colour‑sensing function is integrated onto a miniature chip installed directly beneath the titration cell, delivering direct readout for RGB primary‑colour values.
It achieves substantially higher colour‑measurement sensitivity, facilitates convenient maintenance, and endpoint judgment is unaffected by light‑source positioning or lamp replacement. No calibration or readjustment is required even after disassembly and cleaning of the titration cell.
Technological upgrades for color detectors enable identification of extremely subtle colour transitions, which in turn improves overall titration‑system performance. Near the titration endpoint, titrant volume increases at a very slow rate. This delivers more stable blank values for the Kjeldahl analyzer, superior repeatability for ammonium‑sulfate recovery tests, and higher accuracy for sample measurement results.